Key Features:
Voltage Range: 1-6V
Current Range: 40A, 60A, 80A (800A in parallel mode)
Precision: Up to 1mA with automatic dynamic range switching
Energy Recovery: Optional to the grid for enhanced efficiency
Control Software: BaSyTest Battery Test Software
Applications:
Battery Testing: Ideal for testing and formation of single-cell batteries in research and production environments.
High-Power Cell Testing: Suitable for testing cells in high-current applications like electric vehicles, power tools, and energy storage systems.
Energy Efficiency: The optional energy recovery feature makes the XCTS Mk II particularly well-suited for applications focused on sustainable energy solutions.
Advantages:
High Precision: Achieve up to 1mA precision with automatic and dynamic current range switching.
Compact Design: The small 19” form factor ensures the system fits easily into any lab or production environment.
Scalability: The system can be expanded with parallel operation, reaching up to 800A for high-current applications.
Energy Recovery: Optional grid recovery for efficient energy use and reduced operational costs.
User-Friendly Software: BaSyTest software provides a powerful and intuitive interface for test management.
Technical Data
Specification | XCTS Mk II 40A | XCTS Mk II 60A | XCTS Mk II 80A |
Max. Current per Channel | 40A | 60A | 80A |
Max. Power per Channel | 200W | 300W | 400W |
Parallel Operation | Yes, up to 80A | Optional, up to 600A | Optional, up to 800A |
Range Switching | Automatic and dynamic | Automatic and dynamic | Automatic and dynamic |
Channels per Unit | 8 or 20 | 5, 10, or 12 | 4 or 10 |
Voltage Range | 0-6V | 0-6V | 0-6V |
Min. Discharge Voltage | 1V | 1V | 1V |
Resolution | 0.1mV / 1mA | 0.1mV / 0.05mA | 0.1mV / 1mA |
Accuracy | 1mV / 10mA + 0.1% up to 20A | 1mV / 10mA + 0.1% up to 40A | 1mV / 10mA + 0.1% up to 80A |
20mA + 0.1% up to 40A | 30mA + 0.1% up to 60A | 40mA + 0.1% up to 80A | |
Measurement | Voltage (4W), current, temperature | Voltage (4W), current, temperature | Voltage (4W), current, temperature |
Temperature Input | NTC 5k (Epcos B57861S0502) or GE TK95 | NTC 5k (Epcos B57861S0502) or GE TK95 | NTC 5k (Epcos B57861S0502) or GE TK95 |
Time Resolution / Rise Time | 1µs / 1ms | 1µs / 1ms | 1µs / 1ms |
Sampling Rate | Peak 0.5ms / Min. 10ms | Peak 0.5ms / Min. 10ms | Peak 0.5ms / Min. 10ms |
Current/Voltage Ripple | < 0.01% FS / < 0.5mV | < 0.01% FS / < 0.5mV | < 0.01% FS / < 0.5mV |
Control Mode | Current, voltage, power, resistance, and combinations | Current, voltage, power, resistance, and combinations | Current, voltage, power, resistance, and combinations |
Energy Recovery | Yes, between channels | Yes, between channels | Yes, between channels |
Recovery to the Grid | Optional, total efficiency up to 70% (charge) / 60% (discharge) | Optional, total efficiency up to 70% (charge) / 60% (discharge) | Optional, total efficiency up to 70% (charge) / 60% (discharge) |
Software | BaSyTest | BaSyTest | BaSyTest |
Interface | Ethernet | Ethernet | Ethernet |
Connector | D-SUB 7W2 | D-SUB 9W4 | D-SUB 9W4 |
Noise | 70 – 85dB(A) (without recovery to the grid) | 60 – 83dB(A) (with recovery to the grid) | 60 – 83dB(A) (with recovery to the grid) |
Size/Weight (No Recovery) | 450mm x 310mm x 500mm / 40kg | 450mm x 310mm x 500mm / 40kg | 450mm x 310mm x 500mm / 40kg |
Power Supply | 400VAC / 50Hz / 3 phases + N | 400VAC / 50Hz / 3 phases + N | 400VAC / 50Hz / 3 phases + N |
Size/Weight (With Recovery) | 450mm x 310mm x 620mm / 70kg | 450mm x 310mm x 620mm / 70kg | 450mm x 310mm x 620mm / 70kg |