Magnetic Field Generator and Analyzer MGA 1033
Purpose: To generate and measure magnetic fields in compliance with standards such as MIL-STD-461, ISO 11452-8, and IEC 61000-4-8.
Key Features:
Frequency Range: DC to 250 kHz
Field Strength: Up to 1000 A/m (with triaxial Helmholtz coil)
Modules:
• Signal Generator: DC – 250 kHz
• Power Amplifier: 800 W output power, DC – 1 MHz bandwidth
• Spectrum Analyzer: 16-bit, 1 MS/s sampling rate
Applications:
Automotive and military EMC testing
Avionics and general magnetic field testing
Advantages:
Triaxial Helmholtz Coil Compatibility: Automatically generates fields in all three spatial axes.
Flexible Testing: Supports various standards including MIL-STD-461 and DO-160.
Modular Design: Allows for separate use of signal generator, power amplifier, and spectrum analyzer.
Accessories:
Helmholtz coils
Sensor and field coils
Coupling transformers

Magnetic Field Test System MAG 1000
Purpose: For power frequency magnetic field immunity testing, compliant with IEC 61000-4-8.
Key Features:
Field Strength:
• Continuous: 1 – 120 A/m
• Short-term: 100 – 1100 A/m
• Frequency: 50 Hz / 60 Hz
Test Duration:
• Continuous: 1 min – 8 h
• Short-term: 1 s – 3 s
Dimensions:
• Generator: 550 mm x 450 mm x 143 mm, weight 1.7 kg
• Coil: Approx. 1 m x 1 m, weight 46 kg
Applications:
Testing large devices in both vertical and horizontal planes
Advantages:
Compact and Mobile: Equipped with wheels for easy movement.
User-Friendly: Simple interface and keypad for ease of use.
Robust Construction: Designed for reliable performance in various testing scenarios.
Accessories:
Power supply
Tripod and antenna

Test Setup for Medical Devices
Standards: IEC 60601-1-2 Ed. 4.1 and IEC 61000-4-39
Purpose: To test medical devices for electromagnetic compatibility, ensuring they operate reliably in environments with potential electromagnetic interference.
Key Components:
Magnetic Field Generator:
Frequency Ranges:
• 9 kHz – 150 kHz: MGA RL 120
• 150 kHz – 26 MHz: MGA RL 100-3
• Field Strengths: Up to 1000 A/m
Sensor Coils:
• MGA LS 040: For accurate field strength measurement
• MGA LS 040: For accurate field strength measurement
Matching Networks / Preamplifiers:
• MGA RL 120 NW: For 9 kHz – 150 kHz
• MGA RL 100-3 NW-60601: For 150 kHz – 26 MHz
Test System:
CDG 7000-75-10: RF Generator with 75 watts output power, amplifier (if required), and HELIA 7-MGA software.
Applications:
Medical Devices: Ensuring compliance with IEC 60601-1-2 Ed. 4.1 and related standards
Advantages:
Comprehensive Setup: Covers a broad frequency range with precise measurement capabilities.
Customizable: Preset parameters and limits, with options for custom routines.
Accessories:
Solenoid coils
Brackets and adapters
HELIA 7 software
These setups ensure that your devices meet the necessary standards for electromagnetic compatibility and magnetic field immunity. For further details or specific configurations, consult with our experts to find the best solution for your needs.